Dosimetric Evaluation of 3-Dimensional Conformal Radiotherapy Technique in Postoperative Patients with Gastric Carcinoma: When Is IMRT Really Needed?

  1. Rakesh Kapoor ,
  2. Srinivasa GY ,
  3. Namrata Das ,
  4. Chinna Babu Dracham ,
  5. Divya Khosla ,
  6. Arun S Oinam

Vol 5 No 3 (2020)

DOI 10.31557/apjcc.2020.5.3.151-156

Abstract

Background and purpose: Adjuvant Chemoradiotherapy is the standard of care for postoperative gastric cancers with high risk features. The purpose of the current study is to do a dosimetric analysis in the postoperative setting by using 3-Dimensional Conformal Radiotherapy (3D-CRT) to a total dose of 45 Gy in 25 fractions over 5 weeks. A subsequent comparison with the 3D-CRT and IMRT of other published data is presented.
Materials and Methods: Sixty postoperative patients who underwent radiation with 3D-CRT technique were included in this analysis. All patients received concurrent 5-Flurouracil or Capecitabine chemotherapy along with radiation. Radiation plans were analysed in terms of PTV coverage, conformity index (CI), homogeneity index (HI), organs at risk (OARs) and dose volume histogram (DVH) parameters.
Results: DVH statistics for PTV: Dmean was 45.2±0.8 Gy, D98 was 42.9 ±1 Gy, D95 was 43.4±0.8 Gy, and D2 was 47.7±1.1 Gy. Mean CI for all plans was 1.23±0.43 and HI was 1.09±0.03. DVH parameters for OARs: right kidney, Dmean = 11.9±5.1 Gy, V18 was 21.5±13.8%, V15 was 27.2±14.9% and left kidney, Dmean was 17.7±5.8 Gy, V18 was 33.5±13.8%, V15 was 43.2±15.5%. Dmean for liver was 27.7±6.4 Gy and V20 was 69.2±15.8%. D195cc for the bowel bag was 36.3±10.8 Gy.
Conclusion: The results of this study and subsequent comparison with existing literature suggests that 3D-CRT provides adequate homogenous target volume dose coverage and OAR protection, comparable to IMRT. More than the radiotherapy technique, it was the anastomotic site and the tumor location that determined the OAR doses. 

Introduction

Stomach cancer is the fifth most common cancer in the world (5.7% of all new cases) and the third most common cause of cancer related death (8.2% of all cancer deaths) [1]. India has a low incidence of stomach cancer compared to world statistics and is the fifth most common cancer in India. The estimated incidence of gastric cancer in India for the year 2018 was 5% (57,394 cases) and about 38,818 new gastric cancer cases were estimated to have occurred in males [2]. Partial or total gastrectomy is the mainstay of treatment in the management of stomach cancer [3]. In the majority of cases, treatment with surgery alone is not satisfactory and the 5-years survival depends on tumour stage and lymph nodal status. Loco-regional recurrence is the major problem in 80-85% of patients [4-5]. Since the introduction of SWOG/INT0116 trial in 2001, adjuvant concurrent chemoradiotherapy (CRT) became the standard of care for gastric cancer [6]. The benefits of adjuvant CRT was extended even for patients with D2 dissection [7]. However, post-operative radiotherapy significantly increases toxicities compared to surgery alone [6]. This has been minimised with the advent of newer technologies like 3D-CRT and IMRT. Superiority of IMRT over 3D-CRT has not been clearly established unlike few other sites [8-9]. The purpose of the present study was to do a dosimetric analysis of the doses received in post-operative setting by using 3D-CRT to a total dose of 45 Gy in 25 fractions over five weeks with concurrent chemotherapy. A subsequent comparison with the existing 3D-CRT and IMRT literature was done.

Materials and Methods

This retrospective analysis on dosimetric data in post-operative carcinoma stomach was carried out at the Department of Radiotherapy and Oncology in a multispecialty tertiary care centre in India after informed consent. Being a retrospective dosimetric analysis of patients treated as per institutional protocol, additional ethical clearance was not sought. Case records of all patients with carcinoma stomach (adenocarcinoma and signet ring cell carcinoma) registered in the department between January 2012 and December 2015 were analysed. Out of 156 cases of primary gastric cancer, only 93 (60%) of the cases were histopathologically confirmed to have adenocarcinoma and signet ring cell carcinoma. Out of these, the 60 patients who underwent radical surgery followed by post - operative radiotherapy have been included in this analysis and the rest who received palliative treatment were omitted. All patients received concurrent chemotherapy using 5-FU or capecitabine based regimen. Staging was done according to AJCC 7th edition staging system [10].

Radiotherapy Planning

The post-operative patients of carcinoma of stomach were selected for this dosimetric analysis of the doses received by PTV and OARs using 3D-CRT. All patients underwent total or subtotal gastrectomy with D1 or D2 lymph-nodal dissection. Planning CT scans were taken with patients in supine position using multislice CT scanner with a slice thickness of 2.5mm (GE healthcare technologies, Wankesha, WI, USA). The images were then transferred to EclipseTM treatment planning system (v.8.6, Varian associates, Palo, Alto, CA, USA). Contouring was done as per Tepper and Gunderson guidelines [10-11]. The clinical target volume (CTV) included both post-operative tumour bed and regional lymphatic drainage. The nodal CTV included nodes in the para-aortic (PA), celiac axis, porta hepatis and superior mesenteric artery (SMA) group. All lymph nodal groups were contoured according to the guidelines be Tepper et al [10]. The PA lymph-nodal volume was delineated by giving differential margin to the aorta from thoracic (T11) to lumbar (L2) vertebrae, (i.e. 2.5 cm right side, 2 cm anterior, 1.2 cm left side, and 0.2 cm posterior), 1cm symmetric margin was given to portal vein (from origin of portal vein to porta-hepatis), celiac artery (from origin of celiac axis to its first bifurcation) and superior mesenteric artery (from its origin to 1 cm) (Figure 1).

Figure 1: CTV NODAL Contouring; 1cm Symmetric Margin Around Portal Vein, Celiac Artery, Superior Mesenteric Artery; Differential Margin Around Aorta, 2.5 cm right, 2 cm Anterior, 1.2 cm Left, and 0.2 cm Posterior.

The CTV tumour included the tumour bed, remaining stomach and anastomotic site. Both CTV nodal and CTV tumour were dependent on the primary location of the tumour, pathological T-stage and nodal stage. The CTV total was made by joining both CTV nodal and tumour with the help of Boolean apparatus. PTV was contoured by giving 1 cm symmetric margin around CTV total. OARs including liver, both kidneys, spinal cord, heart, and small bowel were contoured. 3D-CRT plans were generated by using one anterior and two lateral fields with appropriate wedges (Figure 2) with 6 MV photons and treatment was executed by a linear accelerator. A total dose of 45 Gy in 25 fractions were prescribed to the PTV. Plans were optimised whenever required to deliver the prescribed dose to more than 95% of PTV. Dose volume histograms (DVH) were generated and evaluated for all the OARs before delivering the treatment. Evaluation of dosimetric data of the target volumes and OARs was done using Quantitative Analysis of Normal Tissue Effects in Clinics (QUANTEC) parameters [12].

Figure 2: PTV with 95% Dose Colour Wash; 3DCRT Plan Showing Three Radiation Portals, Anterior; Right Lateral, and Left Lateral Fields.

All 3D-CRT plans were analysed in terms of PTV coverage, conformity index (CI), homogeneity index (HI), and OAR dose volume parameters. Several definitions of CI and HI are available but we used the RTOG definitions to calculate the CI and HI.

Conformity index (RTOG definition) = Reference isodose volume / target volume Homogeneity index = maximum isodose in the target / reference isodose

In an ideal scenario the CI should be equal to 1 and HI should be < 2. CI less than 1 indicates that the target volume is not adequately irradiated and a value greater than 1 signifies that the irradiated volume is greater than the target volume.

Statistics

Statistical analysis was done with the Statistical Program for Social Sciences (SPSS v23, IBM Corp, USA). Descriptive analysis was done for the dosimetric and demographic data. Summary of statistics including mean, range, and standard deviation were obtained.

Results

Patient and treatment characteristics

The median age of patients in our study was 52 years (22-75 years). Most of them were males (33 patients, 55%). Most of the patients were smokers (60%) and occasional alcoholics (55%). Five percent of the patients had a family history of malignancy. The common presenting symptoms were loss of appetite, upper abdominal pain and loss of weight. Most of the patients underwent sub-total gastrectomy (90%) with D1 lymph-node dissection (66%) or D2 lymph-node dissection (33%). Adenocarcinoma (intestinal type) was the most common post-operative histopathological variant. All patients underwent post-operative radiotherapy to a total dose of 45 Gy in 25 fractions over 5 weeks using 3D-CRT (Table 1).

Table 1: Patient and Disease Characteristics.

Characteristic Number Percentage (%)
Gender
Male 33 55%
Female 27 45%
Median age (years) 52 (22-75)
Smoker 36 60%
Alcoholics 33 55%
Surgery
Sub-total gastrectomy 54 90%
Total gastrectomy 6 10%
Lymph-node dissection
D1a 40 66%
D2b 20 33%

*D1 lymph node dissection, stations 1 to 6; along the greater and lesser curvature of stomach; bD2 lymph node dissection, D1 and stations 7 to 11; along left gastric artery, common hepatic artery, celiac trunk, and splenic artery.

Dosimetric parameters of PTV

The mean PTV volume was 848.6±337cc, Dmean was 45.2±0.8 Gy, D98 was 42.9 ±1 Gy, D95 was 43.4±0.8Gy,D50 was 45.2±0.7Gy, D5 was 47.3±1.0Gy and D2 was 47.7±1.1Gy. Mean CI for all plans was 1.23±0.43 and HI was 1.09±0.03 (Table 2).

Table 2: Dosimetric Parameters of the Planning Target Volume (PTV).

PTV Mean ± Standard deviation Range
Volumea 848.6 ± 337.0 338 - 2610
Dmeanb 45.2 ± 0.8 41.3 - 47.0
D98b,c 42.9 ± 1.0 39.1 - 44.6
D95b 43.4 ± 0.8 39.5 - 44.8
D50b 45.2 ± 0.7 41.8 - 47.3
D5b 47.3 ± 1.0 44.8 - 49.7
D2b 47.7 ± 1.1 45.6 - 50.7
Confomity index (CI) 1.23 ± 0.43 0.70 - 2.30
Homogenity index(HI) 1.09 ± 0.03 1.04 - 1.20

*PTV volume in cm3; bDose measured in Gray (Gy); cD98 is dose received by 98% of the PTV

Dosimetric parameters of OARs

DVH parameters for OARs; right kidney Dmean was 11.9±5.1 Gy, V18 was 21.5±13.8%, V15 was 27.2±14.9%,V13 was 31.7±15.7% and left kidney Dmean was 17.7±5.8 Gy, V18 was 33.5±13.8%, V15 was 43.2±15.5%, V13 was 59±15.6% (Figure 3).

Figure 3: Dose Volume Histogram Showing V15 for Left Kidney and Right Kidney.

Dmean for liver was 27.7±6.4Gy and V20 was 69.2±15.8%. D195cc for the bowel bag was 36.3±10.8Gy, Dmax was 48.7±1.4Gy, and V45 was 29.2±9.8%. Dmax for spinal cord was 30.0±10.8Gy. Dmean for the heart was 4.5±0.84Gy (Table 3).

Table 3: Dosimetric Parameters of the OARs.

Organ at risk Mean ± Standard deviation Range
Right kidney
Dmeana 11.9 ± 5.1 2.2 - 22.8
V18b 21.5 ± 13.8 0 - 55
V15 27.2 ± 14.9 0 - 61
V13 31.7 ± 15.7 0 - 65
Left kidney
Dmeana 17.7 ± 5.8 8.5 - 45.6
V18b 33.5 ± 13.8 5.7 - 61.0
V15 43.2 ± 15.5 13.5 - 84.0
V13 59.0 ± 15.6 21.6 - 85
Heart
Dmeana 4.5 ± 0.84 3.7 - 8.5
Liver
Dmeana 27.7 ± 6.4 9.0 - 39.9
V20 69.2± 15.8 27 - 95
Bowel Bag
Dmaxa 48.7 ±1.4 44.7 - 52
V45 29.2±9.8 13.5 - 58.8
D195cca 36.3±6.7 23.9 - 48.0
Spinal cord
Dmaxa 30.0 ±10.8 Jul-48

*Dose measured in Gray (Gy); bV18 is volume received by 18Gy in percentage

Discussion

Partial or total gastrectomy along with the D1 or D2 lymph-node dissection is the mainstay of treatment in the management of stomach cancer. Loco-regional recurrence is the major problem in patients treated with surgery alone necessitating various attempts at improving survival using radiotherapy and chemotherapy both before and after surgery [5-11]. Adjuvant CRT became the standard of care since the introduction of SWOG/INT0116 trial [6]. The presence of OARs like spinal cord, liver, small bowel and bilateral kidneys limit the delivery of radiation dose to the post-operative tumour bed, especially with 2D-conformal radiotherapy where antero-posterior and postero-anterior (AP-PA) radiation portals were used. The high rates of Grade 3 toxicities seen in this seminal trial were subsequently minimized with the replacement of conventional treatment techniques by 3D-CRT and IMRT. With the introduction of 3D-CRT, dose to the OARs could be minimized with adequate delivery of dose to the tumour bed and nodal basin [4-12].

In our study, by using the standard RTOG tools, both HI and CI were within the range of reported IMRT data [13]. In a study by Li Z et al, the author compared dosimetric parameters of 5 field intensity-modulated radiotherapy (5F-IMRT), 7F-IMRT, single-arc volumetric modulated arc therapy (SA-VMAT) and double arc (DA)- VMAT [12]. All plans were made with the prescription dose of 50.4 Gy in 28 fractions. CI was 0.86±0.02 in both 5F-IMRT and 7F-IMRT planning’s, 0.83±0.03 for SA-VMAT, and 0.87±0.03 for DA-VMAT planning. CI was better with the IMRT and ARC therapy compared to 3D-CRT planning in the current study due to the forward planning of 3D-CRT.

Kidney is a radiosensitive organ and abdominal or pelvic radiotherapy leads to damage to one or both the kidneys [14]. For bilateral kidneys, V20 <32%, mean dose < 15-18Gy, V12 <55% is recommended to get <5% of radiation induced kidney damage [15-16]. In the present study, mean dose to the left kidney was higher than the right kidney, the primary reason being that the left kidney in most of the patients was located within the target volume and tumor bed was anterior to the left kidney. However, the doses were comparable to that reported in various studies as outlined in Table 4 by both IMRT and 3D-CRT. Therefore, location of the post-operative bed rather than the technique appeared to be the primary limiting factor. Alani et al. defined separate high and low dose kidney instead of laterality to incorporate this concept.

Table 4: Table Showing a Comparison between 3D-CRT and IMRT Plans of Bilateral Kidney Doses.

    Current study (45 Gy) Zhiping et al. (50.4 Gy) [12]     Alani et al. (45 Gy) [9]     Milano et al. (50.4 Gy) [17]     Zhang et al. (45 Gy) [18]     KK Murthy et al. (45 Gy) [19]  
    3D-CRT 5F-IMRT 7F-IMRT 3D-CRT   IMRT 3F   IMRT 3D-CRT   IMRT 4F-CRT   IMRT
Right kidney Dmean 11.9±5.1 12.83±2.15 12.83±2.03 25.6 (High Dose Kidney)   12.8 (High Dose Kidney) 13.45   9.52 11.9±1.4   13.5 ± 0.65 12.0 ± 2.6   12.9 ± 2.5
  V18/V20 21.5±13.8 0.19±0.06 0.23±0.04 39   17       19.2 ± 1.1   16.2 ± 1.1      
Left kidney Dmean 17.7±5.8 15.42±1.93 14.96±1.91 21.2 (Low Dose Kidney)   113 (Low Dose Kidney)     19.35 13.2 ± 1.21   15.3 ± 0.63 13.7 ± 3.1   15.0 ± 2.6
  V18/V20 33.5±13.8 0.27±0.04 0.28±0.04 17   7 23.8     29.9 ± 2.5   27.7 ± 1.8      

Liver is a parallel organ, and the radiation induced liver damage (RILD) is found to correlate with the volume and dose of radiation received by the normal liver tissue. The incidence of RILD is dependent on the mean dose to the normal liver tissue with models predicting no cases of RILD with a mean liver dose < 31 Gy [20]. In the present study, mean dose to the liver was 27.7±6.4 Gy and V20 was 69.2±15.8 which were within the tolerable limit for the normal liver tissue. Though mean doses to the liver were lower with IMRT as reported by Li Z et al [12] and Minn et al. [8], 3D-CRT planning yielded values well within tolerance limits.

Severe gastro-intestinal (GI) toxicities are the main dose limiting factor in case of abdominal irradiation. In a comparison study of IMRT and 3D-CRT for adjuvant treatment of gastric cancer by Minn AY et al, grade 2 or higher GI toxicity was found to be similar between IMRT and 3D-CRT patients (61.2% and 61.5% respectively) and V45 for the bowel space is 106.2 cc, Dmax is 50.5 Gy for IMRT planning [18]. In another study by Liu GF et al, acute GI toxicities were found to be 56% and 54% for IMRT and 3D-CRT respectively [21]. Li Z et al showed that the Dmean for small intestine was 17.19±6.26 Gy for 5F-IMRT, 16.50±5.76 Gy for 7F-IMRT, 18.01±6.34 Gy for SA-VMAT, 16.90±5.92 Gy for DA-VMAT [13].

In the present study, Dmax for the bowel bag is 48.7±1.4 Gy, V45 is 29.2±9.8%, and D195cc is 36.3±6.7 Gy which showed that the doses were comparable to the above mentioned studies.

For individual patients more than the radiotherapy technique it was the position of the anastomotic site which influenced high OAR doses (kidney and liver). The anastomotic sites and DVHs comparing kidney doses of an individual patient are shown in Figure 4.

Figure 4: Position of CTV and OARs.

Limitation of the current study are its retrospective nature and there is no direct comparison of 3D-CRT planning with the IMRT planning for the same set of patients.

In conclusion, results of the study suggest that there is a challenge in achieving the Organ at Risk (OAR) constraints in respect to bowel bag and mean doses achieved for left kidney in patients who had undergone D1 lymphadenectomy. It was also observed that inclusion of postoperative site was having changing position depending upon patient to patient which further led to difficulty in achieving constraints. Some of these patients may favour IMRT based treatment to have better dose delivery.

In the current study we found that post-operative 3D-CRT in cases of carcinoma stomach is acceptable in terms of target volume coverage and homogeneity. The strength of our current study is that it contains detailed dosimetric analysis in terms of physical parameters of target volume, HI, CI and OAR doses in a homogenous group of post-operative cases of carcinoma of stomach. In a country like India where we have increased patient load, we can spend less time on machine using 3D-CRT. Based on the result we could conclude that 3D-CRT offers good coverage of target volumes at the same time sparing the critical surrounding organs.

References


  1. [cited 23 Feb 2020]. Available: https://gco.iarc.fr/today/data/factsheets/cancers/7-Stomach-fact-sheet.pdf .
  2. [cited 23 Feb 2020]. Available: https://gco.iarc.fr/today/data/factsheets/populations/356-india-fact-sheets.pdf .
  3. Adenocarcinoma of the stomach: Review of 1,497 cases Dupont J. Benton, Lee J. Rillens, Burton George R., Cohn Isidore. Cancer.1978;41(3). CrossRef
  4. Post-operative radiochemotherapy for gastric cancer: adoption and adaptation Ringash J., Khaksar S.J., Oza A., Couture J., Japp B., Moore M., et al . Clinical Oncology.2005;17(2). CrossRef
  5. Adenocarcinoma of the stomach: Areas of failure in a re-operation series (second or symptomatic look) clinicopathologic correlation and implications for adjuvant therapy Gunderson Leonard L., Sosin Henry. International Journal of Radiation Oncology*Biology*Physics.1982;8(1). CrossRef
  6. Chemoradiotherapy after Surgery Compared with Surgery Alone for Adenocarcinoma of the Stomach or Gastroesophageal Junction Macdonald John S., Smalley Stephen R., Benedetti Jacqueline, Hundahl Scott A., Estes Norman C., Stemmermann Grant N., et al . New England Journal of Medicine.2001;345(10). CrossRef
  7. An observational study suggesting clinical benefit for adjuvant postoperative chemoradiation in a population of over 500 cases after gastric resection with D2 nodal dissection for adenocarcinoma of the stomach Kim Sung, Lim Do Hoon, Lee Jeeyun, Kang Won Ki, MacDonald John S., Park Chan Hyung, et al . International Journal of Radiation Oncology*Biology*Physics.2005;63(5). CrossRef
  8. Comparison of intensity-modulated radiotherapy and 3-dimensional conformal radiotherapy as adjuvant therapy for gastric cancer Minn A. Yuriko, Hsu Annie, La Trang, Kunz Pamela, Fisher George A., Ford James M., et al . Cancer.2010;116(16). CrossRef
  9. Limited Advantages of Intensity-Modulated Radiotherapy Over 3D Conformal Radiation Therapy in the Adjuvant Management of Gastric Cancer Alani Shlomo, Soyfer Viacheslav, Strauss Natan, Schifter Dan, Corn Benjamin W.. International Journal of Radiation Oncology*Biology*Physics.2009;74(2). CrossRef
  10. The American Joint Committee on Cancer: the 7th Edition of the AJCC Cancer Staging Manual and the Future of TNM Edge Stephen B., Compton Carolyn C.. Annals of Surgical Oncology.2010;17(6). CrossRef
  11. Adjuvant therapy for gastric cancer: Current and future directions Foo Marcus. World Journal of Gastroenterology.2014;20(38). CrossRef
  12. Dosimetric comparison of intensity modulated and volumetric arc radiation therapy for gastric cancer LI ZHIPING, ZENG JIANSHUANG, WANG ZI, ZHU HONG, WEI YUQUAN. Oncology Letters.2014;8(4). CrossRef
  13. Radiation therapy oncology group: Radiosurgery quality assurance guidelines Shaw Edward, Kline Robert, Gillin Michael, Souhami Luis, Hirschfeld Alan, Dinapoli Robert, Martin Linda. International Journal of Radiation Oncology*Biology*Physics.1993;27(5). CrossRef
  14. Clinical radiation nephropathy Cassady J.Robert. International Journal of Radiation Oncology*Biology*Physics.1995;31(5). CrossRef
  15. Use of Normal Tissue Complication Probability Models in the Clinic Marks Lawrence B., Yorke Ellen D., Jackson Andrew, Ten Haken Randall K., Constine Louis S., Eisbruch Avraham, et al . International Journal of Radiation Oncology*Biology*Physics.2010;76(3). CrossRef
  16. Quantitative Analyses of Normal Tissue Effects in the Clinic (QUANTEC): An Introduction to the Scientific Issues Bentzen Søren M., Constine Louis S., Deasy Joseph O., Eisbruch Avi, Jackson Andrew, Marks Lawrence B., et al . International Journal of Radiation Oncology*Biology*Physics.2010;76(3). CrossRef
  17. Intensity-modulated radiation therapy in the treatment of gastric cancer: early clinical outcome and dosimetric comparison with conventional techniques Milano M T, Garofalo M C, Chmura S J, Farrey K, Rash C, Heimann R, Jani A B. The British Journal of Radiology.2006;79(942). CrossRef
  18. Double-arc volumetric modulated therapy improves dose distribution compared to static gantry IMRT and 3D conformal radiotherapy for adjuvant therapy of gastric cancer Zhang Tao, Liang Zhi-Wen, Han Jun, Bi Jian-Ping, Yang Zhi-Yong, Ma Hong. Radiation Oncology.2015;10(1). CrossRef
  19. Evaluation of dose coverage to target volume and normal tissue sparing in the adjuvant radiotherapy of gastric cancers: 3D-CRT compared with dynamic IMRT Murthy KK, Shukeili KA, Kumar SS, Davis CA, Chandran RR, Namrata S. Biomedical Imaging and Intervention Journal.2010;6(3). CrossRef
  20. Analysis of radiation-induced liver disease using the Lyman NTCP model Dawson Laura A., Normolle Daniel, Balter James M., McGinn Cornelius J., Lawrence Theodore S., Ten Haken Randall K.. International Journal of Radiation Oncology*Biology*Physics.2002;53(4). CrossRef
  21. Clinical Outcomes for Gastric Cancer following Adjuvant Chemoradiation Utilizing Intensity Modulated versus Three-Dimensional Conformal Radiotherapy Liu Gene-Fu F., Bair Ryan J., Bair Eric, Liauw Stanley L., Koshy Matthew. PLoS ONE.2014;9(1). CrossRef

Author Details

Rakesh Kapoor
Professor

Srinivasa GY
Assistant Professor
drsrinivasagowda@gmail.com

Namrata Das
Affiliation not stated

Chinna Babu Dracham
Affiliation not stated

Divya Khosla
Affiliation not stated

Arun S Oinam
Affiliation not stated

How to Cite

Kapoor, R., GY, S., Das, N., Dracham, C. B., Khosla, D., & Oinam, A. (2020). Dosimetric Evaluation of 3-Dimensional Conformal Radiotherapy Technique in Postoperative Patients with Gastric Carcinoma: When Is IMRT Really Needed?. Asian Pacific Journal of Cancer Care, 5(3), 151-156. https://doi.org/10.31557/apjcc.2020.5.3.151-156
  • Abstract viewed - 0 times
  • PDF (FULL TEXT) downloaded - 0 times
  • XML downloaded - 0 times